I am processing HL7 message. OBX-4field gives containment hierarchy.I see different dotted hierarchy in the message. is there any standard for containment hierarchy dotted number?
For below example. will first dotted 1 always mean MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS and 2nd dot 5-always mean MDC_DEV_ECG_VMD. are these number configurable in the medical device.I want to store data uniquely based using MDS/VMD/CHAN.
Right now I am getting HL7 from one source..will these hierarchy will always be same for that source. is this would be valid if i get hl7 message from other source.
MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS/MDC_DEV_ECG_VMD/MDC_ECG_HEART_RATE to Acehive
1.5.0.1
1-MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS
5-MDC_DEV_ECG_VMD
13-MDC_DEV_METER_PRESS_BLD_VMD
1-MDC_DEV_METER_PRESS_BLD_CHAN
OBX|1||69965^MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS^MDC|1.0.0.0|||||||X
OBX|2||69798^MDC_DEV_ECG_VMD^MDC|1.5.0.0|||||||X
OBX|3|NM|147842^MDC_ECG_HEART_RATE^MDC|1.5.0.1|88|{beat}/min^{beat}/min^UCUM|||||R|||20200508051804.8340+0530||||DFG~01^^Y71A57FFFE6188F3^EUI-64
OBX|4|NM|148066^MDC_ECG_V_P_C_RATE^MDC|1.5.0.2|7|{beat}/min^{beat}/min^UCUM|||||R|||20200508051804.8340+0530||||DFG~01^^Y71A57FFFE6188F3^EUI-64
OBX|17||69854^MDC_DEV_METER_PRESS_BLD_VMD^MDC|1.13.0.0|||||||X
OBX|18||69855^MDC_DEV_METER_PRESS_BLD_CHAN^MDC|1.13.1.0|||||||X OBX|19|NM|150018^MDC_PRESS_BLD_DIA^MDC|1.13.1.15|68|mm[Hg]^mm[Hg]^UCUM|||||X|||20200508051804.8340+0530||||DFG~01^^Y71A57FFFE6188F3^EUI-64||unknow
So far after exploration i understood that those dotted numbers are not standard.Below is IHE documents. Probably it depends on the profile. which can be different at different point.
https://www.hl7.org/documentcenter/public/wg/healthcaredevices/N0262_WG7_RCH.1g.pdf
I marking it is as answer. I will change this if something better explanation comes.